65
Alarms
Parameter
Low Limit, Default
High Limit, Default
Resolution
IBP 2 - Systolic
Neonatal
-9.9 to 94.9, 15
-4.9 to 99.9, 50
0.1mmHg
Paediatric
-9.9 to 94.9, 15
-4.9 to 99.9, 50
0.1mmHg
Adult
-9.9 to 94.9, 15
-4.9 to 99.9, 50
0.1mmHg
IBP 2 - Diastolic
Neonatal
-9.9 to 94.9, 5
-4.9 to 99.9, 20
0.1mmHg
Paediatric
-9.9 to 94.9, 5
-4.9 to 99.9, 20
0.1mmHg
Adult
-9.9 to 94.9, 5
-4.9 to 99.9, 20
0.1mmHg
IBP 2 - MAP
Neonatal
-9.9 to 94.9, 10
-4.9 to 99.9, 25
0.1mmHg
Paediatric
-9.9 to 94.9, 10
-4.9 to 99.9, 25
0.1mmHg
Adult
-9.9 to 94.9, 10
-4.9 to 99.9, 25
0.1mmHg
etCO
2
(mmHg)
Neonatal
0 to 84, 25
15 to 99, 60
1mmHg
Paediatric
0 to 84, 25
15 to 99, 60
1mmHg
Adult
0 to 84, 25
15 to 99, 60
1mmHg
etCO
2
(kPa)
Neonatal
0 to 11.2, 3.3
2 to 13.2, 8
0.1kPa
Paediatric
0 to 11.2, 3.3
2 to 13.2, 8
0.1kPa
Adult
o to 11.2, 3.3
2 to 13.2, 8
0.1kPa
etCO
2
(%)
Neonatal
0 to 11, 3.2
1.9 to 13, 7.8
0.1%
Paediatric
0 to 11, 3.2
1.9 to 13, 7.8
0.1%
Adult
0 to 11, 3.2
1.9 to 13, 7.8
0.1%
Содержание Smartsigns Compact SC1000
Страница 1: ...SC1000 Smartsigns Compact...
Страница 11: ...11 Accessories Module 2 NIBP 1 x NIBP Hose NIBP NIBP 3 x NIBP Cuffs...
Страница 82: ......