40
SpO
2
Monitoring
14.4 SpO
2
Error Messages
The following section provides the user with a description of ERROR
messages, which may be displayed during respiration
measurement.
Error Message
Cause
Rectification
'NO SENSOR'
No sensor connected to
system
Connect the sensor to
the system
'SENSOR OFF
PATIENT'
Sensor detached from
patient
Re-apply sensor to the
patient
'SENSOR
DEFECTIVE'
System detected a
defective sensor
Replace sensor
'LOW
PERFUSION'
Patient perfusion is too
low
Check patient status and
patient connection
'PULSE SEARCH'
Hardware is being
adjusted for best signal
Change site or sensor
'TOO MUCH
AMBIENT LIGHT'
Excessive light for the
sensor to function
Reduce the ambient
level around the patient
'INSUFFICIENT
SIGNAL'
Patient perfusion is too
low
Check patient status and
patient connection