49
Invasive Blood P
ressure Monitoring
Connect the transducer cable to the transducer paying particular
attention to the channel identification (I or II).
17.2 Priming the system
Follow the manufacturers priming
instruction.
Actuate the flush system gently and
continue to prime without pressure.
Pressurise the system to 300 mmHg and
fast flush the system for 2 -3 seconds.
Check for air bubbles.
The system is now ready to zero and
calibrate
Operating the
FLUSH system
Each of the free ends must be inserted into the appropriate blue
socket/s on the IBP module labelled I and II.