32
www.humancaregroup.com
Altair
Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
Emission test
Compliance
Electro magnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Human Care lift uses RF energy only for its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The Human Care lift is suitable for use in all establishments including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC
61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electro magnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Electrical fast trans
-
ient / Burst
IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power
supply lines
n/a. for input/output
lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
+/- 1 kV differential
mode
n/a. for common mode
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT))
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Human Care lift requires continued ope
-
ration during power mains interruptions, it is recommended that the lift
be powered from an uninterruptible power supply or battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment
NOTE UT is the a.c. mains voltage prior to application of the test level.