Guidance and manufacturer’s declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electro magnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80MHz to 2,5GHz
10 V/m
800MHz to 2,5GHz
3 Vrms
3 V/m
10V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the Human Care lifts, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = [3,5/3]
P
d = [3,5/3]
P
80 MHz to 800 MHz
d = [7/3]
P
800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance
level in each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following symbol.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Human Care lifts is are
used exceeds the applicable RF compliance level above, the Human Care lifts should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Human Care lifts.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the Human Care lifts.
The Human Care lifts are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the Human Care lifts can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communi
-
cations equipment (transmitters) and the Human Care lifts as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2√P
80 MHz to 800 MHz
d = 1,2√P
800 MHz to 2.5 GHz
d = 0,7√P
0.01
0.12
0.12
0.07
0.1
0,37
0,37
0,22
1
1,16
1,16
0,7
10
3,67
3,67
2,21
100
11.6
11.6
7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Recycling
Once the product has reached the end of its life,
it should be disposed according to your coun-
try’s rules, especially for the separate collection
of electrical and electronic products.
Correct disposal helps prevent negative
consequences for the environment and human
health.
Содержание Altair 55200H
Страница 82: ...82 www humancaregroup com Altair...
Страница 83: ...www humancaregroup com 83 DEUTSCH...