7
The setup menu is used by users for SET Stroke, IGN. mode, Fuel type, test mode
and time setup (refer to the 5th item below for time setup).
The choice of Stroke and IGN.M. mode will affect the test of speed. Their relation is:
Stroke
4 Stroke
2 Stroke
IGN.M. mode
Once
Double
Once
Double
RPM calculation
×2
×1
×1
÷2
Therefore if Stroke or IGN.M. mode were wrongly chosen, RPM calculation would have
error of integral times.
If choosing “gasoline”, test result of HC adopts hexane normal (‘HC’ is chosen for display);
if choosing LPG, test result of HC adopts propane normal (C
3
H
8
is chosen for display).
Test mode can either be set here or reset in test menu.
Continuously press Key-S to choose in sequence
Stroke
, IGN. mode
,
Fuel type
,
Test mode
,
Time setup
and
Return
. The former 4 Items
can be reset by pressing Key-▲ or Key-▼, after chosen via Key-S.
When cursor appears at
Return
,
press Key-E for exiting from SET and returning to
main menu image.
3.6 ADJ and Inspection
Choose
ADJ
in main menu via Key-S and then press Key-E, then the following adjust
Setup menu
hh : mm
:
ss
Stroke
:
√
4 stroke
2 stroke
IGN. mode
:
√
once
double
Fuel type
:
√
gasoline
LPG
Test mode
:
Idle
Time setup
resturn
S:select, E:confirm,∧∨ modify