VEGA78
EN - 32
Fig. 59: Analyzer Configuration screen for Two phase system 3-wire 3P OY – ENG type
Fig. 60: Analyzer Configuration screen for 2-phase system 3-wire 3P OY – ENG_UK type
1. Move the cursor using arrows keys on field relative to “
System
”, marked with blue
background.
2. Use
F3
or
F4
keys (alternatively touch
MOD(+)
o
MOD(-)
) to select the type of system
choosing among “4-WIRE”, “3-WIRE”, “ARON” or “SINGLE” options.
3. Press
SAVE
or
ENTER
keys (or the smart icon
) to save the selected option and
confirm by “Ok”. This selected parameter will be present also after turning off the meter.
4. Press
ESC
key (or the smart icon ) to quit without saving any modification.
5.3.1.1. Setting system frequency
1. Move the cursor using arrows keys on field relative to “
Freq[Hz]
”, marked with blue
background.
2. Use
F3
or
F4
keys (alternatively touch
MOD(+)
o
MOD(-)
) to select the system
frequency choosing between
50Hz
or
60Hz
options. This parameter is relevant ONLY if
the input voltage not permits to recognize the frequency value (for example, only the
clamps for the current measurement are connected). In this case the instrument
generates an internal synchronism equal to the value of the set frequency.
3. Press
SAVE
or
ENTER
keys (or the smart icon
) to save the selected option and
confirm by “Ok”. This selected parameter will be present also after turn off the meter.
4. Press
ESC
key (or the smart icon ) to quit without saving any modification.