6.5.5. Anomalous results
1. Display of message "
not
rdy
" indicates that IMP57 is not reacting the commands sent
by FULLTEST4050 through serial interface. Check that FULLTEST4050 is connected
to IMP57 by means of C232NG1 cable and that IMP57 is powered (green LED).
2. If after pressing
START
key, the message “
180
Err.
” is displayed and a 3 second long
beep sound is emitted, it means that the voltage detected by IMP57 does not reach the
minimum preset limit.
3. If after pressing
START
key, the message “
480
Err.
” is displayed and a 3 second long
beep sound is emitted, it means that the voltage detected by IMP57 exceeds the preset
limit.
4. If after pressing
START
key, the message “
HOT
” is displayed and a 3 second long
beep sound is emitted, it means that the temperature probe reached a too high
temperature value.
5. If after pressing
START
key, the message “
Curr.
Err.
” Is displayed and a 3 second
long beep sound is emitted, it means that the streaming current is lower than 10A.
6. If after pressing
START
key, the message “
Err.
5
” is displayed and a 3 second long
beep sound is emitted, it means that the instrument detects a mistake in data coding by
IMP57.
7. If after pressing
START
key, the message “
O.r.
” is displayed and a long beep sound is
emitted, breaking off just by pressing
STOP
key, it means that the measured
impedance is higher than the maximum impedance value.