IMP57
EN - 28
6.4.
OPERATING INSTRUMENT FOR “TYPE 4” INSTRUMENT
Operating Instruction are given according to
Table 1 - Instrument Type
definition.
6.4.1. Instrument Settings
1.
Press power switch.
2.
Press the
FUNC
key to select
CONTINUITY
.
3.
Press the
MODE
key to select
EXT
mode. The instrument
shows a screen as nearby.
m
A
V
4.
Press the
SET DISPLAY
key to
select the mode of setting the
magnetothermal protection
nominal current
In
. Through the
arrow keys
select the nominal current.
A
5.
Press the
SET DISPLAY
key
once again to select the mode
of setting the protenction type.
Through the arrow keys
select the protection type (type
B or C).
Press the
SAVE
key to confirm the modifications.
6. OR
Press the
ESC RECALL
key to quit without saving.