2–8
ESL9000 Series Tape Library Unpacking and Relocating Guide
Selecting an Installation Location
Environmental Conditions
The installation site must provide the following environmental conditions during
normal library operation:
•
Humidity: 20% to 80% non-condensing
•
Temperature: 59°F to 90°F (10° to 32°C)
•
Altitude: sea level to 10,000 feet (3,048 m)
NOTE:
For additional library specifications refer to Appendix A of the
hp StorageWorks
ESL9000 Series Tape Library Reference Guide.
Physical Distance Between
the Library and a Host Workstation
The maximum physical distance between the library (SCSI bulkhead) and the SCSI
connector on a host workstation is dictated by the Low Voltage Differential (LVD) bus
characteristics (see Table 2-1 and Table 2-2).
NOTE:
For the ESL9198, ESL9322, and most ESL9326/ESL9595 libraries, a SCSI bus
extender is used for bus 1 (robotic controller, drive 0, drive 1.) For the other drive pairs (D2, D3),
(D4, D5), (D6, D7), (D8, D9) no internal SCSI extender is used. See Table 2-1 and Table 2-2 for
the maximum external SCSI bus cable lengths.
SCSI bus extenders can be purchased for other drive pairs.
Table 2–1: Physical Distances (ESL9198/ESL9322)
Bus
Devices
Maximum External Length
1
Robot, D0, D1
20 meters
2
D2, D3
10 meters
3
D4, D5
10 meters
4
D6, D7
10 meters