NOTE
For best measurement accuracy, perform an HP 3458A multimeter ACAL
not more than 24 hours prior to these adjustments. Each DAC channel to
be adjusted MUST be in the non-calibrated mode (CALx:STAT OFF), or
errors in the adjustment constants will occur.
Equipment Setup
Connect a 1 k
Ω
resistor and HP 3458A multimeter to the channel to be
adjusted, as shown in figure 5-1. Be sure to set all three channel jumpers to
the V (voltage) position. Set the HP 3458A to DCV function and set NPLC
100.
NOTE
The value of 1 k
Ω
in Figure 5-1 is the factory value used for original
adjustments. If you previously adjusted your DAC with a different
resistance value, use that resistance value rather than 1 k
Ω
.
Figure 5-5. Voltage Adjustment Connections
38 Adjustments
Chapter