USING AUTOTRANSFORMERS. If the mainframe is to be
energized via an autotransformer (for voltage reduction) make
sure the common terminal is connected to neutral (that is, the
grounded side of the main’s supply).
WARNING
CAPACITOR VOLTAGES. Capacitors inside the mainframe may
remain charged even when the mainframe has been
disconnected from its source of supply.
USE PROPER FUSES. For continued protection against fire
hazard, replace the line fuse(s) only with fuses of the same
current rating and type (such as normal blow, time delay, etc.).
Do not use repaired fuses or short-circuited fuseholders.
CAUTION
MAXIMUM VOLTAGE. Maximum voltage that may be applied
between any two terminals within the same channel is
±
15 Vdc. Do not
apply voltage between any pair of terminals if the D/A Converter is
turned off.
STATIC ELECTRICITY. Static electricity is a major cause of
component failure. To prevent damage to the electrical components in
the DAC, observe anti-static techniques when removing a DAC from
the mainframe or when working on the DAC. Also, when installing a
DAC in a mainframe slot, be sure to tighten the front panel screws.
POWER-UP/DOWN OUTPUTS. At power-up or down, potentials up
to 12V may appear across the channel outputs for up to 10 msec.
Chapter 1
General Information 11