Chapter 5
Adjustments
Introduction
This chapter contains procedures to adjust the HP E1328A D/A Converter
(DAC), including voltage adjustments and current adjustments.
The DAC should be adjusted after repair to ensure peak performance.
Equipment required for the adjustment procedures is listed in Table 1-2,
Recommended Test Equipment.
Before performing adjustments, the DAC must have a minimum 60-minute
warm up and the line voltage must be 115/230 Vac
±
10%. For best
accuracy, the test area temperature should be between 18°C and 28°C, and
stable to
±
1°C.
WARNING
Do not perform any of the following adjustments unless you are
a qualified, service trained person and have read the
WARNINGS and CAUTIONS in Chapter 1.
Voltage
Adjustments
This adjustment is used to update the stored voltage constants for each DAC
channel, and should be perfomed:
•
On initial setup
•
Periodically (24 hours or 90 days) to maintain specified output accuracy
(see Appendix A - Specifications of the HP E1328A User’s Manual)
•
When temperature has changed more than
±
5°C from last adjustment
temperature
•
When the type of load is changed
•
Any time accuracy is in doubt