![HP 64700 series Скачать руководство пользователя страница 141](http://html.mh-extra.com/html/hp/64700-series/64700-series_user-manual_161481141.webp)
Selecting the Sample
Period or Rate
You can select the sample period (or rate, if it’s more convenient)
with the "sample" command. Valid periods in the standard mode
are between 10 ns (100 MHz) and 50 ms (20 Hz). Valid periods in
the glitch capture mode are between 20 ns (50 MHz) and 50 ms (20
Hz). The accuracy of the sample rate is that of the crystal oscillator,
approxi /- 0.01% .
The command
sample period_is 10 nsec <RETURN>
sets the sample rate to once every 10 nanoseconds.
1
Specifying the
Trigger Condition
The "mode" and "sample" commands control the acquisition of
data, but they do not tell the analyzer when to acquire data. This is
done by using the "trigger" command to specifying a trigger
condition.
The trigger condition is the combined specifications of all labels
(that is, the pattern, edge, or glitch specifications for each label
OR ed together to form the complete specification for the 16
external data signals).
The "trigger" command is part of the trace specification.
Trigger on Anything
The default condition is to trigger on any activity on the probe
signals.
The command
trigger on anything <RETURN>
sets the analyzer to its default condition.
Trigger on Pattern
In this type of trigger condition, a specific set of signal activities, a
pattern, must be present in order for the trigger condition to be
met.
You specify a data pattern consisting of 1s, 0s, or X’s (don’t cares)
on one or more of the labels or label signals. The most significant
bit is probe signal 15, or the highest bit number for the label
Timing: Using the Analyzer 8-7
Содержание 64700 series
Страница 2: ......
Страница 19: ...Tables Table 9 1 Summary of Commands 9 4 Table 9 2 Command Assignments 9 5 Contents 11 ...
Страница 20: ...1 Notes 12 Contents ...
Страница 24: ...1 Notes 1 4 Introduction ...
Страница 106: ...1 Notes 4 36 Performance Measurements ...
Страница 118: ...Figure 5 8 True Demultiplexing 5 12 Using the External Analyzer ...
Страница 128: ...1 Notes 6 6 Timing Introduction ...
Страница 134: ...1 Notes 7 6 Timing Getting Started ...
Страница 168: ...1 Notes 8 34 Timing Using the Analyzer ...
Страница 176: ...1 compare This command specifies a post processing compare definition Syntax 9 8 Timing Commands ...
Страница 182: ...1 copy This command copies specifications displays or measurement data to selected output Syntax 9 14 Timing Commands ...
Страница 191: ...define DATA xbits 8 width 8 RETURN Related Commands delete rename threshold Timing Commands 9 23 ...
Страница 200: ...Examples execute repetitively RETURN Related Commands halt halt_repetitive_execution 9 32 Timing Commands ...
Страница 201: ...1 find This command finds a trigger like event in trace memory Syntax Timing Commands 9 33 ...
Страница 219: ...1 mark This command marks specified conditions in trace memory Syntax Timing Commands 9 51 ...
Страница 236: ...1 process_for_data This command limits the trace list display to specified samples Syntax 9 68 Timing Commands ...
Страница 254: ...1 trigger This command specifies trigger conditions Syntax 9 86 Timing Commands ...
Страница 261: ...Examples display timing_diagram RETURN waveform_size medium RETURN Related Commands present Timing Commands 9 93 ...
Страница 262: ...1 Notes 9 94 Timing Commands ...
Страница 280: ...1 Notes B 16 Timing Diagrams and Outputs ...
Страница 292: ...1 Notes C 12 Timing Messages ...