6. Troubleshooting
92
Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
EMPTY CONTAINER
N
A/V
The pump has delivered
the programmed
container size or the
pump has delivered the
programmed VTBI; AND
Auto KVO is delivering
(if available in current
program).
1
Press [
SILENCE
] to
quiet the alarm for 2
minutes.
2
Press [
STOP
] to place
the pump in the
STOP
mode and clear the
alarm.
3
Use the
NEW
CONTAINER
function to
repeat the current
program. Refer to
Starting a New
Container
in Chapter 2
Basic Operation.
Note: Titration is allowed at
this time.
FLASHING DISPLAY
•
Line four of the
display flashes.
N
A/V
A Non-programming
Callback has occurred
because the pump has
been waiting for more
than three minutes for a
keypress on a non-
programming input or a
confirmation display.
Follow the instructions on the
display to clear the alarm.
Alert Display Message Is the
pump
infusing?
Alar
m
Cause/Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
430-10984-001.book Page 92 Friday, March 31, 2006 4:49 PM
Содержание GEMSTAR 13088-04-01
Страница 61: ...Notes 54 Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 54 Friday March 31 2006 4 49 PM...
Страница 105: ...Notes 98 Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 98 Friday March 31 2006 4 49 PM...
Страница 115: ...Notes 108 Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 108 Friday March 31 2006 4 49 PM...
Страница 133: ...Notes 126 Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 126 Friday March 31 2006 4 49 PM...
Страница 151: ...Notes 144 Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 144 Friday March 31 2006 4 49 PM...
Страница 161: ...Hospira GemStar Manual 430 10984 001 Rev 03 06 430 10984 001 book Page 156 Friday March 31 2006 4 49 PM...