5 FUNCTION
87
5.5.3
Perform the SAMPLING RESET
1.
Press the SAMPLING RESET button on the SAMPLING SETTING screen.
A screen confirming whether or not you want to reset the sampling time is displayed.
2.
Press [YES].
Returns to the SAMPLING SETTING screen.
The lock is released, and sample water will be obtained again.
Displaying the SAMPLING SETTING screen" (page 85).
MEAS. screen
MAINTENANCE screen
SETTING1/2 screen
[SETTING]
[MAINTE.]
[SAMPLING SETTING]
SAMPLING SETTING
[RESET:]