5 FUNCTION
110
5.13 Alarm
You can activate an alarm when exceeding the specified measurement values for each
measurement item.
Reference
The operations and settings for alarms are as follows.
Item
Description
Checking an alarm
(
You can check the current alarm.
Cancelling an alarm
(
Forcibly cancels an activated alarm.
Checking the alarm
history
(
A maximum 511 alarms can be viewed.
Deleting the alarm
history
(
Deletes the alarm history.
ALARM TIME
(
An alarm occurs when a specified value is exceeded for more than a specified period of time.
Also, an active alarm will be cancelled if the measurement value settles within a specified
value for more than a specified period of time. However, when performing the washing and the
calibration operation, a counter is reset.
HIGH ALARM
(
When exceeding the specified measurement value for each measurement item, an alarm
indicating an abnormal sample water is activated.*
(Alarm No. 10, 20, 30, 40, 50, 60, 70)
LOW ALARM
(
When falling short of the specified measurement value for each measurement item, an alarm
indicating an abnormal sample water is activated.*
(Alarm No. 11, 21, 31, 41, 51, 61, 71)
UNIT HIGH ALARM
(
When exceeding the specified measurement value for each measurement item, an alarm
indicating an abnormal device status is activated.*
(Alarm No. 12, 22, 32, 42, 52, 62, 72)
UNIT LOW ALARM
(
When falling short of a specified measurement value for each measurement item, an alarm
indicating an abnormal device status is activated.*
(Alarm No. 13, 23, 33, 43, 53, 63, 73)
*:
Since this device rounds measured and specified values down to the significant digits,
an alarm may not be activated even if it is displaying a measured value with the value
equal to the value specified for the alarm.