4 CALIBRATION
55
5.
Remove the conductivity sensor from the residual chlorine/conductivity cell.
6.
Completely dry the electric conductivity sensor using paper, such as filter paper.
Note
Be careful not to scratch the sensor.
7.
Turn ON the power.
8.
Press the ZERO VALUE button ([x.x]) on the TURB. CAL. screen.
The zero calibration value input screen is displayed.
Conductivity sensor
Enlarged
view
Sample water feed valve must
be closed.
Electric-conductivity sensor connector
Installation location
[x.x]