
4 CALIBRATION
36
4.
Enter the zero calibration value using the numerical key pad and press [SET].
Input range: 0.0 CU to 20.0 CU
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COL. CAL. screen.
5.
Open the valve of the calibration tank.
6.
Press [ZERO].
A confirmation screen is displayed asking whether to start calibration.
7.
Press [YES] to start zero calibration.
The screen shows that the calibration is in progress.
8.
Adjust the flow rate to 50 mL/min using the flowmeter valve.
Wait until the calibration is completed (10 minutes).
Tip
[FIX]: Forcibly stops the calibration and updates it without waiting for automatic completion
[STOP]: Stops the calibration
Mark showing the 80 mL/min position
50 mL/min position (three gradations below 80 mL/min position)
Flow rate reading position (top surface of the float)
Valve
Содержание TW-100
Страница 1: ...Water Distribution Monitor TW 100 Instruction Manual CODE GZ0000080187H ...
Страница 8: ......
Страница 142: ...7 MAINTENANCE 130 ADC Connectors on the printed board Residual chlorine sensor Conductivity sensor ...
Страница 174: ......