Calibration
33
7.
If needed, execute purging (extraction, cell filling, and drainage) referring to "Screen
operations in the manual measurement/calibration mode" (page 14).
8.
Execute extraction and fill the cell at any time referring to steps 3. to 6. and "Screen
operations in the manual measurement/calibration mode" (page 14).
9.
Press the ENT button.
Measurement starts. The operation guide display shows [ESC: Cancel].
The elapsed time is shown (counting up). When measurement ends, the calibration result
appears.
Fig. 33 Manual span calibration - Measurement
When measurement ends normally, the measured value is held.
The operation guide display shows [ESC: Return].
Fig. 34 Manual span calibration - Result
10.
Press the ESC button.
The instantaneous value measurement screen returns.
11.
Execute drainage referring to "Screen operations in the manual measurement/calibration
mode" (page 14) at any time.
12.
Press the ESC button.
The start state of the manual span calibration mode returns.