Measurement
56
Span calibration
Perform span calibration using the conditions in the table below.
Note
Do not introduce water during span calibration
Also set [Span Point] in the calibration settings to the same value as the net concentration of the span
liquid, because water is not used.
Reference
"Calibration Setting screen" (page 80)
"Span liquid for calibration" (page 21)
Measurement
Perform measurement using the conditions in the table below.
Reference
"Measurement Setting screen" (page 70)
Parameter
Condition
Span Point
Concentration of span liquid
Extraction Time
0 sec
Separation Time
0 sec
Number of Purge
3 or more
Span liquid for calibration
Use 24 mL of span liquid for calibration.
Parameter
Condition
Extraction Time
0 sec
Separation Time
0 sec
Number of Purge
3 or more
Measurement liquid
Take 24 mL from the measurement liquid.