Basic Operation
14
Fig. 17 USB icon
Screen operations in the manual measurement/calibration mode
On the instantaneous value measurement screen, which appears when the manual
measurement mode, manual zero calibration mode, or manual span calibration mode starts,
you can switch to any operations below to control extraction, cell filling, measurement, and
drain manually.
Stirring motor: Rotate/stop the stirring blades inside the extraction tank.
Fill cell valve: Start/stop flow from the extraction tank to the measurement cell.
Drainage valve: Start/stop drainage from the measurement cell and extraction tank.
Reference
"Manual measurement" (page 45)
"Manual zero calibration" (page 30)
"Manual span calibration" (page 32)
The procedure for performing operations with the instantaneous value measurement screen
appearing is explained below.
Fig. 18 Example of instantaneous value measurement screen
(manual measurement)
The buttons and button functions, which can be used with the instantaneous value
measurement screen appearing, are described in the table below.
Table 1 Operable buttons with the instantaneous value measurement screen
Button
Function
ENT button
Starts measurement.
Up button
Opens and closes the fill cell valve (UPPER V).
Down button
Opens and closes the drainage valve (LOWER V).
Right button
Switches the stirring motor ON and OFF (STIR).
ESC button
Returns to the mode start state.
Содержание OCMA-500
Страница 1: ...Oil Content Analyzer OCMA 500 Instruction Manual CODE GZ0000331784D ...
Страница 20: ......
Страница 58: ...Measurement 38 Flow scheme of operation Fig 36 Operation flow of automatic measurement ...
Страница 65: ...Measurement 45 Manual measurement Flow scheme of operation Fig 45 Operation flow of manual measurement ...
Страница 140: ......