19
Electrode Preparation
Calibration
Measurement
After Measurement
C
alibration operation
Measure standard solutions in ascending order of ion concentration to create
calibration curves in the meter (
“Measurement principle of ion selective
electrode” (page 56)).
Perform calibration once every day or before measurement as appropriate.
Before operation
Be sure to fill with internal solution and rinse the electrode
before the calibration operation. (
solution” (page 12), “Rinsing the electrode” (page 15))
Perform conditioning before the first use, before use after
storage for two weeks or longer, or if the electrode is less
responsive.(
Note on page 23)
Prepare at least two types of standard
solution with different concentrations.
Always use new standard solution for
calibration.
After the standard solutions (50 mL) and
ISA (50 mL) included in the package are
used up, use the standard solution (500
mL) and ISA (500 mL) available as
consumables.
Note
Note
Syringe
Standard
solutions
Items to be prepared
ISA (50mL)
Note
Note