68
S
pecifications
The following table shows the specifications of the product.
Table 8 Specifications of combination type potassium ion selective electrode
*1: The measurement accuracy is the difference between the measured value
and the assigned value of the 100 mg/L standard solution after the two
points calibration using the 10 mg/L and the 100 mg/L standard solutions.
The temperature in the calibration and in the measurement is 25°C.
*2: The ion concentration value enclosed in ( ) is the measurable pH tolerance
range.
However, the tolerance range differs depending on the ion concentration.
*3: This is the electrode potential gradient in the calibration using the 10 mg/L
and the 100 mg/L standard solutions at 25°C.
Item
Description
Operating temperature range
0°C to 50°C
Internal electrode
Ag/AgCl (silver/silver chloride) electrode
Internal solution
1 mol/L NaCl
Liquid junction
Movable sleeve type
Wetted part (part contacting
sample solution) material
PVC, PSU
Measurement range
0.39 mg/L to 39000 mg/L K
+
(1
10
-5
mol/L to 1 mol/L K
+
)
Measurement accuracy (*1)
±2%
pH range (*2)
pH 5 to pH 11 (3.9 mg/L K
+
)
Potential gradient (*3)
53 mV/decade to 62 mV/decade
Insulation resistance
5
10
10
Ω
Portion to be placed
From ion selective membrane to internal
solution fill port
Connector and cable length BNC,
1
m
Содержание 6582S-10C
Страница 1: ...Instruction Manual Combination type Potassium Ion Selective Electrode 6582S 10C ELECTRODE...
Страница 18: ...8...
Страница 52: ...42...
Страница 62: ...52...
Страница 79: ...69 Specifications External dimensions Unit mm 16 5 O D 18 85 150...