30
15
Shake the electrode to remove any air bubbles remaining
on the ion selective membrane.
yyy
16
Perform the second calibration.
Wait for a stable reading and perform the second calibration as described
in the Meter Instruction Manual.
When the second calibration is finished, the slope (%) of the electrode is
displayed.
17
Check the slope (%) on the meter.
For details on how to check the slope (%), refer to the Meter Instruction
Manual.
It is recommended to place the electrode in the electrode stand.
If air bubbles remain on the ion selective
membrane, it may result in inaccurate
measurement.
Tip
Tip
Shake the
electrode
View video
View video
Note
Note