Troubleshooting
Has the composition
and concentration of
a standard solution
changed?
If the standard solution has
expired, replace it with a new one.
Use up the standard solution
within approximately 3 months
after opening the bottle.
Furthermore, do not return the
standard solution to its original
container after it has been used
once.
(
―
Is the amount of the
discharging internal
solution appropriate?
Press the upper gray part of the
electrode cap.
If the internal solution is
discharged too much, it may
cause measurement errors or a
decrease in the response rate
during calibration.
Check item
Solution
View in video