Make sure that the inside of the electrode moves up and
down and that internal solution oozes out from the liquid
junction.
7
Close the rubber stopper of the
internal solution fill port.
It is best that the internal solution oozes out slightly. If the
outflow is extremely little, the potential of the reference electrode
will become unstable and susceptible to stir.
In such a case, repeat step 5 and 6 until the internal solution
oozes out from the liquid junction.
The internal solution
oozes out from the
liquid junction.
View video
View video
Tip
Tip