S
CALE
O
PERATION
S
ECURITY
S
EAL
I
NSTALLATION
Wire Security Seal (Seal Conversion Kit
PN
46-00359)*
1. Temporarily remove the platter and place it in a safe location.
2. Install the calibration button sealing cover and through-hole sealing screw.
3. Tighten the screw and route the wire seal through the holes in the plate and screw.
Figure 60. Calibration Switch/Button Wire Security Seal Assembly
4. Leaving the least amount of slack in the wire loop as possible, close the wire seal and verify the seal connection is secure.
5. Reinstall the platter.
Figure 61. Platter Installation
*
These procedures are for MS2320 StratosH Scanner/
Diva
Scale models only
.