S
CALE
O
PERATION
C
ALIBRATION
Bar Code Calibration Procedure without Remote Display*
Unit Calibration (Kilograms)
For Pound weight units see instructions on page 45.
Kilograms Calibration
³
9
9
6
0
4
4
Figure 51. Entering Bar Code Calibration Mode (Kilograms)
If a razz tone sounds, an error has occurred. Refer to
Diagnostic Indicator Display; Error Codes
starting on page 33 for
additional information.
6. Make sure there is
no load
on the scale platter. The white LEDs will be used to indicate the current step in the calibration process.
7.
Calibrate a zero load.
The white LEDs will blink once periodically. Wait 8 to 10 seconds for scale stability, and then press the
Multi-Function Button once. After the Multi-Function Button has been pressed, there will be a short delay followed by one short
beep indicating that the zero load value has been stored.
Figure 52.
*
These procedures are for MS2320 StratosH Scanner/
Diva
Scale models only
.
46