43-TV-25-30 Iss.6 GLO Aug 07 UK
Reference Temperature: 22°C
Reference Sample Rate: 2Hz (500msec)
Reference Humidity: 65% RH +/-15%
Long term stability: 0.2%/year
* Does not includes reference junction calibration of ±1.0 º C using the standard “ice bath” method of calibration. Factory accuracy can be im-
proved by performing a field calibration. Also does not include any error on the sensor.
** Tolerance for these input types includes that of the external shunt resistors (0.1% tolerance)
*** Reference Accuracy can be improved to +/- 0.4ºC/0.7ºF using the single point compensation calibration.
# not available on the
QXe.
Including all Field Cal values.
PT100
α
= 0.00385
-328 to 1562
-200 to 850
1.1
0.6
0.01%/ °C
PT200
α
= 0.00385
-328 to 1562
-200 to 850
1.1
0.6
0.01%/ °C
PT500
α
= 0.00385
-328 to 1562
-200 to 850
1.1
0.6
0.01%/ °C
PT1000
α
= 0.00385
-328 to 1562
-200 to 850
1.1
0.6
0.01%/ °C
100 ohm Nickel
-76 to 356
-60 to 180
0.9
0.5
0.01%/ °C
120 ohm Nickel
-112 to 500
-80 to 260
0.5
0.3
0.01%/ °C
Cu10#
-328 to 500
-200 to 260
5.5***
3***
0.01%/ °C
Cu53#
32 to 302
0 to 150
0.5
0.3
0.01%/ °C
Specification Options
Pulse Input (optional)
(not available on the
QXe
)
4 isolated inputs per board, frequency – 1Hz to 25kHz, updated once per sec. Input: Low < 1V, High >4V
to <50V or Volt free input: Low = short circuit, High = open circuit.
Alarm Outputs (optional)
Programmable alarm set points (6 per pen) can be configured to activate up to 16 outputs for the
QX
and
48 outputs for the
SX
. Update rate: 200 ms for all alarms. Number/Type:
• 4 or 8 relay contacts SPDT, 3A 240VAC, 3A 24VAC/DC, 0.2A 240VDC (non-inductive, internally
suppressed)
• 8 I/O or 16 I/O - SPNO 1A 24VDC (non-inductive, internally suppressed).
16 I/O not available for
the
QXe
Activation: Fully programmable internal alarm levels. Assignable to any relay or discrete output.
Analogue Outputs
(Re-transmission Outputs)
(optional)
(not available for the
QXe)
2 or 4 re-transmission outputs available for the
QX
Recorder and 2, 4, 6 or 8 re-transmission outputs avail-
able for the
SX
Recorder; a pen drives each output. Analog inputs, totalised values or any mathematical
result can be re-transmitted.
Update Rate: 250 msec all channels Accuracy: ±0.1% 0-500
Ω
load, +/-0.25% 500
Ω
1K
Ω
load
Type: 0 to 20 / 4 to 20 mA Maximum Load Resistance: 1000 Ohms
Resolution: 0.002% Isolation: 300VAC
Digital Input/Output
(optional)
8 I/O or 16 I/O: all channels may be selected freely as either digital inputs or outputs. The Digital I/O card
also has 4 channels that can be set as pulse inputs (channels 1 to 4). The operating frequency for pulse
inputs on the Digital I/O card is 1kHz max. Input: Low < 1V, High >4.5V to <10V DC (9V to 20V p-p AC) or
Volt free input: Low = short circuit, High = open circuit. 16 I/O not available for the
QXe
4 relay outputs: all four channels are relay outputs only.
8 relays/ 2 DI card: two outputs can be configured for use as digital inputs: A digital input is provided by a
volt free contact between the normally open (NO) and the common (C) terminals of an output relay. If the
2 Digital inputs are used only 6 relay outputs are available. Closed <500 ohms, Open >300 kohms.
Transmitter Power (optional)
QX
has 200mA @ 24VDC ± 3VDC.
SX
has 1 Amp @ 24VDC ± 3VDC.
QXe
is available with 130mA @ 24VDC TXP as an option on the Comms card.
Agency Approval (optional)
CSA (Optional) CSA22.2-No.1010.1-2004 Certificate Number 211230. UL (Optional) ANSI/UL61010-1-
2004 File # 201698. FM Class 1 Division 2 (optional). FM not available for the
QXe
Vutronik Connection
(optional) QX Recorder only
Optional rear cover with 50-pin connector for direct connection to the recorder. TDC2000/3000 system
using Vutronik Trend Recorder, 24VDC instrument power only.
Extended Security System
(ESS) (optional)
Provides full support for 21 CFR Part 11. Includes features for entry of unique User ID’s and associated
passwords, timeout on inactivity (1 to 10 min.), password expiration (1 to 365 days), up to 50 users, pass-
word re-entry lock out for incorrect entry of password more than 3 times, no re-use of passwords (pro-
grammable 4 to 12 times), traceability by user name
Miscellaneous
Optional customer ID Tagging (3 lines of up to 22 characters each line)
Содержание Minitrend QX
Страница 2: ...ii 43 TV 25 30 Iss 6 GLO Aug 07 UK User manual ...
Страница 50: ...44 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 192: ...186 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 210: ...204 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 242: ...236 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 245: ...43 TV 25 30 Iss 6 GLO Aug 07 UK 239 Section 10 Spares List Minitrend QX Recorder 16 5 3 1 4 2 24 10 49 9 45 19 ...
Страница 249: ...43 TV 25 30 Iss 6 GLO Aug 07 UK 243 Multitrend SX Recorder 14 15 32 46 50 30 31 35 36 28 55 ...
Страница 250: ...244 43 TV 25 30 Iss 6 GLO Aug 07 UK 52 11 23 33 39 5 6 RS485 USB host Ethernet 1 2 3 4 ...
Страница 254: ...248 43 TV 25 30 Iss 6 GLO Aug 07 UK eZtrend QXe Recorder 16 5 3 1 4 2 24 49 9 45 19 6 49 ...
Страница 258: ...252 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 274: ...268 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 276: ...270 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 296: ...290 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 308: ...302 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 312: ...306 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Страница 323: ......