Messages and Diagnostic Codes
Page 70
34-ST-25-20 MC Toolkit User Manual Release
3
9/06
HART MESSAGES
Error code 12: URV too low
URV too low or out of range.
Error code 13: Parameter out of limits
Parameter is out of the valid range.
Error code 14: Span Too Small
Span Too Small
Error Code 14: Warning: New LRV pushed
URV over Sensor Limit
The new URV has caused the URV to go over the sensor limits.
Error code 16: Access restricted
Access to the parameter is not allowed.
Error code 32: Device is Busy!
Device is currently busy; try command again.
Error Code 64: Command not implemented in
the Transmitter!
Action is not supported by the transmitter.
Error on Burst Response!
The burst message contained a communications error.
HART Communication Error!
MC Toolkit detected a communications error in the HART
message.
Invalid command number from the Transmitter.
Please make sure that another secondary
master is not on the network!
The HART protocol supports only one secondary master on the
network.
Manufacturer Code not be found
Manufacturer Code not be found
No device found!
No HART devices could be found on the network.
Non primary variable out of limits
Process applied to the non-primary variable is outside the
operating limits of the field device.
Parity Error!
Bit 6 on first response code byte set. MC Toolkit detected a
communications error in the HART message.
Overrun Error!
Bit 5 on first response code byte set
Framing Error!
Bit 4 on first response code byte set. MC Toolkit detected a
communications error in the HART message.
Checksum Error!
Bit 3 on first response code byte set
Reserved Field Error!
Bit 2 on first response code byte set
Receive Buffer
Overflow!
Bit 1 on first response code byte set
Primary variable out of limits
Process applied to the primary variable is outside the operating
limits of the field device.
Undefined Response code for the command
associated with the current operation!
Response code (First byte) value 1 is returned from the
transmitter. This response value is undefined in the currently
supported HART devices