50461/GTEX MMC/1111
UTImeter Gtex MMC
43
10.9 Verification and adjustment of
the reading index
To verify or to adjust the reading index, in
particular after having renewed a tape, apply the
following instruction:
-
if the equipment is fitted with a 2" connector
(Q2) remove the clip and the collar as shown
on Figure 10-12;
-
unlock and remove the lid of the tape housing;
-
put the tape cleaner on "DOWN" position;
-
keep the equipment standing vertically on a
flat surface;
-
gently lower the tape until the sensor touches
the surface (Figure 10-12);
-
adjust the index to the value corresponding to
the connector Q1 or Q2, as shown on Figure
10-12;
-
In case of a 2" connector (Q2) put back the
clip and the collar .
-
Put the housing lid back and lock it.
IMPORTANT NOTE: these adjusting values for
the reading index are different from the heights
shown in the section 8 "Examples of installation of
the gauging system". They take into account the
recessment of the reaction point from the sensor
tip end and other mechanical parameters.
Figure 10-12
UNFORTUNATELY THE MMC DESIGN / TUBES ARE
NOT COMPATIBLE WITH THE CONCEPT OF
VERIFICATION AND ADJUSTMENT OF THE
READING INDEX DESCRIBED IN THIS SECTION.
THEREFORE PLEASE SEND YOUR EQUIPMENT TO
ONE OF OUR OFFICIAL SERVICE STATIONS FOR
THIS ADJUSTMENT.