GasAlertMicro 5/PID/IR
User Manual
54
Alarm Setpoints
8. When all of the sensor due dates have been set or bypassed,
the alarm setpoints need to be set or bypassed. The following
screen displays.
Factory alarm setpoints may vary by region. Refer to
Resetting
Gas Alarm Setpoints
for an example.
Alarms can be set to any value within the detection range of
the selected sensor. Refer to “Detection Range” in
Specifica-
tions
.
Note
To disable an alarm setpoint, set it to
0
(zero).
If a new setpoint is entered but not confirmed within 5 seconds
by pressing
C
, the following screen displays.
The previous setpoint is retained and the detector
proceeds to the next setpoint.
The setpoints are set in the following order:
•
TWA (if applicable)
•
STEL (if applicable)
•
low
•
high
To bypass a setpoint, press
C
to save the current value and
proceed to the next setpoint.