To maintain the detector in good operating condition, perform the
following basic maintenance as required.
• Calibrate, bump test, and inspect the detector at regular intervals.
• Maintain an operations log of all maintenance, calibrations, bump
tests, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents,
soaps, or polishes. Refer to
Sensor Poisons and Contaminants
.
• Do not immerse the detector in liquids.
Battery Cautions
a
Warning
To avoid personal injury and/or property damage, adhere
to the following:
• Replace the batteries immediately when the detector emits a
low battery alarm.
• Use only batteries that are recommended by BW Technologies
by Honeywell. Refer to
Specifications
.
• Ensure the alkaline batteries are properly installed in the
detector battery pack.
• Charge batteries using only a recommended BW charger. Do
not use any other charger. Failure to adhere to this caution
can cause fire and/or explosion.
• The detector must be deactivated to charge the battery.
• Do not calibrate the detector during or immediately after
charging.
• Both the rechargeable lithium battery pack and the alkaline
battery pack are user-changeable in hazardous locations, but
the alkaline battery cells inside the pack can only be replaced in
a safe area that is free of hazardous gas.
•
Warning:
The battery pack (M5-BAT08) that is equipped with a
lithium polymer cell may present a risk of fire or chemical burn
hazard if misused. Do not recharge, disassemble, heat above
212ºF (100ºC), or incinerate.
•
Warning:
Do not use any other lithium batteries with the Micro5/
PID/IR detectors. Use of any other cell can cause fire and/or
explosion. To order a replacement lithium battery pack (M5-
BAT08), refer to
Replacement Parts and Accessories
.
•
Warning:
Lithium Polymer cells exposed to heat at 266ºF
(130ºC) for 10 minutes can cause fire and/or explosion.
• Dispose of used lithium cells immediately. Do not disassemble
and do not dispose of in fire. Do not mix with the solid waste
stream. Spent batteries must be disposed of by a qualified
recycler or hazardous materials handler.
• Keep lithium cells away from children.
Charging the Battery
To charge the battery, refer to the
GasAlertMicro 5/PID/IR Charger
Instruction Sheet.
Содержание GasAlertMicro 5IR
Страница 1: ......
Страница 8: ...GasAlertMicro 5 PID IR User Manual Title Page ...
Страница 10: ...GasAlertMicro 5 PID IR User Manual ...
Страница 12: ...GasAlertMicro 5 PID IR User Manual ...
Страница 104: ...GasAlertMicro 5 PID IR User Manual ...
Страница 105: ...iERP 128900 D5615 5 English BW Technologies 2009 All rights reserved ...