Activating the Detector
If no passcode is entered, or it is entered incorrectly, the follow-
ing screen displays.
To enable/disable this option, refer to
Due-Lock
in the user
options menu. Also refer to
Passcode Protect
.
Force Calibration Enabled
12. If
Force cal
(force calibration) is enabled in tech mode, calibra-
tion is mandatory before the detector enters normal operation.
Refer to
Force Calibration
in
Tech Mode
to enable/disable, and
refer to
Calibration Procedure
to calibrate.
If
C
is not pressed to start calibration, the following screen dis-
plays and the detector deactivates.
Bump Daily Enabled
a
Caution
BW recommends that a bump test to all sensors be
performed every 24 hours prior to the beginning of the
work shift.
13. If
Bmp Daily
(bump daily) is enabled in tech mode, the follow-
ing screens display.
If a bump test of the LEL and O
2
sensor is not performed, the
detector will deactivate.
Apply the LEL gas and then apply the O
2
(a higher or lower
percentage than the default 20.9%, such as 18% O
2
).