GasAlertMicro 5/PID/IR
User Manual
30
Automatic Oxygen (O
2
) Calibration
Logger Option
The
Logger
option is used to define how often the detector records a
datalog sample (once every 1 to 127 seconds).
From the user options menu, press
H
to scroll to
Logger
. Press
C
within 20 seconds to confirm.
Press
G
or
H
to change the current logger rate. When the desired
value displays, press
C
within 5 seconds to confirm the new value.
When the
Autocal
option is enabled, it
forces the detector to automatically
calibrate the oxygen sensor during
startup.
If the
Autocal
option is enabled, ensure
the detector is activated in a safe area
that is free of hazardous gas in an
atmosphere of 20.9% oxygen.
From the
Sensor
menu, press
H
to scroll
to
O
2
and press
C
within 20 seconds to
confirm.
Press
H
to scroll to
Autocal
. Press
C
to toggle between enable/disable.
The detector is shipped with the
Autocal
option enabled.