Pump Alarm
The pump draws air over the sensors continually. If the pump stops operating or becomes
blocked, the detector activates the pump alarm.
During a pump alarm, HIGH and
display, and
flashes.
Note: The pump alarm displays the same as a high alarm, except that
flashes. Verify that a
pump alarm is occurring before clearing the hose.
The pump alarm continues until the alarm is acknowledged by pressing
.
1.
Clear the blockage and press
to acknowledge. If Force Block Test is enabled,
the detector automatically launches a pump test to verify the pump is operating
correctly.
2.
If the pump test is successful, the detector returns to normal operation.
If the pump test is unsuccessful, refer to See "Troubleshooting" on page
more information. for possible causes and solutions.
BW MaxXT II
65
User Manual
Содержание BW MaxXT II
Страница 1: ...BW MaxXT II 1 2 3 and 4 Gas Detector ...
Страница 2: ......
Страница 56: ......
Страница 68: ......
Страница 74: ...Connecting the Gas Cylinder to the Detector BW MaxXT II 73 User Manual ...
Страница 102: ......
Страница 108: ...P N 129541 L3 EN Rev E 2021 October BW MaxXT II 107 User Manual ...