If Lockout on Self-Test Error is enabled, the sensor must be replaced to activate and
operate the detector. Refer to See "Replacing a Sensor or Sensor Filter" on page
Lockout on Self-Test Error Option Disabled: If this option is disabled and a sensor fails
the self-test, the LCD displays an error message for the applicable sensor. The startup
tests continue.
Note: Enabled sensors are continually tested while the detector is operational.
9.
Automatic Zero for H2S, CO, O2, and LEL
To initiate auto zero during startup, the Auto Zero on Startup option must be enabled
(sensors are enabled individually).
If the Force Calibration option is enabled and the sensor(s) is past due for calibration,
the sensor will not auto zero during startup. The sensor must be calibrated.
Note: If the Auto Zero on Startup option is not enabled in Safety Suite Device
Configurator (SSDC) for any of the sensors, this startup test is bypassed. If ambient air is
set to be measured as 20.8% vol., the automatic oxygen calibration screen displays
20.8% instead of 20.9%.
BW MaxXT II
21
User Manual