3.
Enter the gas concentration value in the Calibration Gas (%) field for O
2
and LEL.
4.
Click Save
Bump Interval
Note: Honeywell recommends to bump test the sensors before each day’s use to confirm their
ability and response to gas by exposing the detector to a gas concentration that exceeds the high
alarm setpoints. Verify that the audible and visual alarms activate. Calibrate if the readings are not
within the specified limits.
Define how often a bump test must be performed for each sensor in the Bump Interval (days)
field. A different bump interval can be set for each sensor.
1.
Enter the value (1-365 days) for each sensor.
2.
Enter 0 to disable the bump interval option. Entering 0 automatically deactivates the
Force Bump user option.
The detector is shipped with the Force Interval option disabled.
BW MaxXT II
46
User Manual