PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD Display
shows
abnormal
result
The cuff position was not correct or it
was not properly tightened
Apply the cuff correctly and try again
Body posture was not correct during
testing
Review the “BODY POSTURE DURING
MEASUREMENT” sections of the
instructions and re-test.
Speaking, arm or body movement, angry,
excited or nervous during testing
Re-test when calm and without
speaking or moving during the test
Irregular heartbeat (arrhythmia)
It is inappropriate for peoplewith
serious arrhythmia to use this Electronic
Sphygmomanometer
9. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS
The following guidelines for assessing high blood pressure (without regard to age or gender) have been
established by the World Health Organization (WHO). Please note that other factors (e.g. diabetes, obesity,
smoking, etc.) need to be taken into consideration. Consult with your physician for accurate assessment, and
never change your treatment by yourself.
Classification of blood pressure for adults
10. TECHNICAL ALARM DESCRIPTION
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if the determined blood pressure
(systolic or diastolic) is outside the rated range specified in part SPECIFICATIONS. In this case, you should
consult a physician or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or
inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear
automatically after about 8 seconds.
LCD shows low
battery symbol
Low battery
Change the batteries
LCD shows
“Er 0”
Pressure system is unstable before
measurement
Don’t move and try again.
LCD shows
“Er 1”
Fail to detect systolic pressure
LCD shows
“Er 2”
Fail to detect diastolic pressure
LCD shows
“Er 3”
Pneumatic system blocked or cuff is too
tight during inflation
Apply the cuff correctly and try again
LCD shows
“Er 4”
Pneumatic system leakage or cuff is too
loose during inflation
LCD shows
“Er 5”
Cuff pressure above 300mmHg
Measure again after five minutes. If the
monitor is still abnormal, please contact
the local distributor or the factory.
LCD shows
“Er 6”
More than 3 minutes with cuff pressure
above 15 mmHg
LCD shows
“Er 7”
EEPROM accessing error
LCD shows
“Er 8”
Device parameter checking error
LCD shows
“Er A”
Pressure sensor parameter error
11. TROUBLESHOOTING