12
G
B
power supply
Battery powered mode: 6Vdc 4*AAA alkaline batteries
Display mode
digital lcd V.A.128*50mm
Measurement mode
oscillographic testing mode
Measurement range
rated cuff pressure:
0kPa–40kPa (0mmHg~300mmHg)
Measurement pressure: 5.3kPa–30.7kPa
(40mmHg–230mmHg)
pulse value: (40–199) beat/minute
accuracy
Pressure:
5ºc–40ºc within ± 0.4kPa (3mmHg)
pulse value: ±5%
normal working condition
temperature: 5ºc to 40ºc
relative humidity ≤85%rH
Atmospheric pressure: 86kPa–106kPa
Storage & transportation condition
temperature: -20ºc~60ºc
relative humidity: 10%~93%
Atmospheric pressure: 50kPa–106kPa
Measurement perimeter of the
upper arm
About 22cm~42cm
Weight
Approx. 320g (Excluding the batteries)
External dimensions
Approx. 182*100*39mm
attachment
4*AAA alkaline batteries, manual, cuff
Mode of operation
continuous operation
Degree of protection
type BF applied part
protection against ingress of water
iP21
Software version
vA3
Specifications
Warning: no modification of this equipment is allowed.