TLi
IQ
System User Manual
7 – 5
Section 7 – Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
11
Liquid control
failed.
Liquid Control
Review the control procedures and repeat the test.
Verify the control has not expired, and is neither cloudy
nor discolored.
If the control fails a second time, call technical service.
12
Invalid liquid control
test result.
Analyzer Internal
Controls: Analyzer
Fail/Cassette Pass
Analyzer Internal
Controls: Analyzer
Pass/Cassette Fail
Refer to Error/Invalid Code table.
Run TLi
IQ
QCette to verify analyzer is functioning properly.
Do not bump or jar analyzer during the test.
Review the Rapid fFN Cassette Kit directional insert to
ensure the correct procedure was followed.
Examine the cassette. Cassette imperfection may cause
an invalid test result. Rerun control on a fresh cassette.