Section 3 – General Operating/Testing Instructions
TEST PATIENT
(Internal Incubation Mode)
1. From the Main Menu, press 1 to select TEST PATIENT.
2. Enter USER ID and press ENTER.
3. Enter the
last two digits
of the CASSETTE LOT# (on cassette
pouch) and press ENTER. The lot number must be entered to
proceed to the next step.
NOTE:
The analyzer will automatically compare the CASSETTE LOT#
used to set calibration with the cassette lot number used for
patient testing. If the lot numbers do not match, the analyzer will
request the user to recalibrate the system. The cassette lot
number used for calibration will be displayed on the third line.
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC
↓
TEST PATIENT
USER ID
>JOHN SMITH
ENTER-CONFIRM
TEST PATIENT
CASSETTE LOT#
>L1002
ENTER-ACCEPT