TLi
IQ
System User Manual
3
−
3
Section 3 – General Operating/Testing Instructions
4. Enter the CALIBRATION CODE# (on cassette box label) and press
ENTER. The code number must be entered to proceed to the next
step.
NOTE:
The calibration code is established by Hologic for each lot of
Rapid fFN Cassettes.
5. When calibration is complete, the system will automatically display
and print the result if AUTOPRINT is set to ON, or it may be
printed/reprinted by pressing the PRINT/ENTER key.
6. Press ESC to return to the Main Menu.
FETAL FIBRONECTIN
03:00 PM 10/26/07
SYSTEM CALIBRATED
USER:JOHN SMITH
→
SET CALIBRATION
CALIBRATION CODE#
>FG56
ENTER-ACCEPT