Section 2 – Installation
FACTORY DEFAULT SETTINGS
The TLi
IQ
System uses the following default settings. To customize the unit to your laboratory requirements, refer to
Section 4, Software Functions – Detailed Descriptions.
The Default Settings are as follows:
AUTOPRINT
Factory setting is Autoprint ON. After every result, the printer will generate a printed result.
INCUBATION MODE
Factory setting is INTERNAL Mode. The incubation mode refers to the timing of the incubation process and the initiation
of the cassette analysis.
In the INTERNAL mode, the analyzer will time the incubation and automatically start the analysis when the incubation is
complete.
In the EXTERNAL mode, the user will be responsible for timing the incubation and for starting the analysis.
NOTE: The INTERNAL mode is recommended by the manufacturer to ensure proper timing of the assay.