The analyzer automatically compares the CASSETTE LOT# used to set
calibration with the cassette lot number used for testing controls. If the lot
numbers do not match, the test process cannot continue. When this
occurs, the cassette lot number used for calibration will be displayed,
and the user is prompted to recalibrate the system.
From the CONTROL TEST MENU, select 1-NEGATIVE CONTROL or 2-
POSITIVE CONTROL.
The most recent CONTROL LOT# is always displayed. Press ENTER to
accept the lot number, or enter a new control lot number. This field will
accept up to 12 alphanumeric characters.
CONTROL TEST MENU
1-NEGATIVE CONTROL
2-POSITIVE CONTROL
NEGATIVE CTL LOT#
>M1023
ENTER-CONFIRM
CASSETTE LOT CHANGED
CALIBRATE SYSTEM
L1002
ESC-MAIN MENU