7 – 1
Section 7 – Troubleshooting
SECTION 7
−
TROUBLESHOOTING
GENERAL INFORMATION
The TLi
IQ
analyzer software is designed for easy troubleshooting. Always heed the beep tones and follow the display
screen prompts to obtain the best performance from your System. The following table lists potential problems, sources of
trouble, and recommended solutions. Call Hologic Inc. Technical Service for any questions related to the performance of
your TLi
IQ
System.
ITEM
PROBLEM
SOURCE
SOLUTION
1
Analyzer display
screen is blank.
Analyzer Power Cord
and Adapter
ON/OFF Switch
Ensure analyzer power cord is firmly connected to
analyzer.
Ensure analyzer power adapter is plugged into a
grounded AC electrical outlet.
Ensure analyzer ON/OFF switch is in ON position.
2
Error Code is
displayed when
analyzer is first
turned on.
Analyzer
Turn analyzer off and back on to reinitialize the system.
If the Error Code persists, refer to Error/Invalid Code table.