1. Apply Sample ID bar code to sample vials.
Place vials in trays for processing.
2. Power on the ThinPrep 3000 Processor, if not
already on.
Wait for System Ready main menu screen.
3. Perform Lube O-Rings maintenance (only
needed at the start of a shift).
4. Check/load supplies:
• Filter trays (if a partial tray is present, do
not remove it)
• Slide cartridges
• Empty staining racks
• Sample vial trays
5. Perform Fixative Spray Preventative
Maintenance if the instrument has been idle for
more than an hour. See Daily Maintenance
Section.
6. Close all doors.
7. Press
Start
Batch
.
• If the instrument detects missing or depleted
supplies, read the message on the display and
load or replenish those items. Close all doors
and press
Continue
to begin processing.
• If the instrument prompts for mainte-
nance, read the message on the display
and replace or empty the item(s) listed.
Close all doors and press
Continue
to
begin processing.
8. When the
Batch
Complete
screen appears at
the end of processing, open the instrument
doors. Carefully remove staining racks and
sample vial tray(s). Tear off the batch report
and match report(s) to trays A and B. Identify
any samples requiring reprocessing or further
attention.
The ThinPrep
®
3000 Processor Quick Reference Guide
Slide Printer Ribbon
Staining Rack
CellFyx Solution
Sample Vials
Slide Waste Bin
Filter Waste Box
Microscope Slides
Slide Cartridge
ThinPre
p
Microscope
Slides
00001
168
!
MAX
MAX
CellFyx
SOLUTION
ª
Waste Bottle
Results Printer Paper
Sample Vial Trays
Filter Trays
Test Vial
(for Maintenance)
Routine Operation
© 2017, Hologic, Inc.
Part No. AW-03369-001 Rev. 002
Daily Maintenance
Lubricate O-Rings
Lubricate the inner surface of the pneumatic test vial. Load
it into tray position A1 of an empty sample vial tray. From
the Maintenance menu, select
Lube O-Rings
.
Fixative Spray Preventative Maintenance
When the fixative menu screen appears, toggle the fixative
dispense lever 3-4 times and press
Spray
. Repeat as needed.
Empty Slide Waste Bin
Remove the slide waste bin. If slides are present, check the
print quality to see if the printer ribbon prints darkly
enough. Dispose of any slides. Replace the slide waste bin.
Check Filter Waste Bin
Look in the filter waste bin to see if it needs to be emptied.
Dispose of filters according to your laboratory guidelines.
Instrument-Prompted Maintenance
Pneumatic System Testing
Lubricate the inside surface of the pneumatic test vial. Load
the test vial into tray position A1 of an empty sample vial
tray. From the test menu, select
Pneumatic Test
.
Clean the Slide Path
Clean the slide translator, slide printer area, slide cell transfer
arm, fixative spray area, staining rack area.
Empty Waste Bottle
Detach the waste bottle from the tubing connections, cap and
remove. Dispose of liquid waste according to your laboratory
guidelines. Lubricate the waste cap o-ring; reconnect the
waste bottle.
Add Fixative Fluid / Insert Clean Fixative Shield
Fill the fixative reservoir with a bottle of CellFyx™ Solution.
Remove the plastic fixative shield and dispose of the removable
cylinder. Replace it with a new one. From the Fixative
Maintenance screen, press
Level
to recalibrate the fixative level.
Slide Printer Ribbon
Replace printer cartridge when prompted. Always be sure the
ribbon is seated in the ribbon guide properly.
Refer to the ThinPrep 3000 Processor Operator's Manual for complete operating instructions.
Содержание ThinPrep 3000
Страница 1: ......
Страница 4: ...The ThinPrep Processor The ThinPrep Processor ...
Страница 5: ...MAN 03939 001 Rev 004 page 1 of 13 Instructions for Use ...
Страница 18: ...Table of Contents Table of Contents ...
Страница 23: ...1 Introduction 1 Introduction ...
Страница 41: ...1 INTRODUCTION 1 18 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 42: ...2 ThinPrep 3000 Installation 2 ThinPrep 3000 Installation ...
Страница 54: ...2 THINPREP 3000 INSTALLATION 2 12 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 55: ...3 PreservCyt and CellFyx Solutions 3 PreservCyt and CellFyx Solutions ...
Страница 71: ...4 Sample Collection and Preparation 4 Sample Collection and Preparation ...
Страница 80: ...5 Instrument Operation 5 Instrument Operation ...
Страница 105: ...6 Maintenance 6 Maintenance ...
Страница 147: ...6 INSTRUMENT MAINTENANCE 6 42 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 148: ...7 Troubleshooting 7 Troubleshooting ...
Страница 204: ...7 TROUBLESHOOTING 7 56 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 205: ...8 Staining and Coverslipping 8 Staining and Coverslipping ...
Страница 211: ...8 STAINING AND COVERSLIPPING 8 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 212: ...9 The ThinPrep Pap Test Training Program 9 The ThinPrep Pap Test Training Program ...
Страница 215: ...10 User Interface Screens 10 User Interface Screens ...
Страница 226: ...Index Index ...
Страница 232: ...INDEX Index 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 233: ...Service Information Service Information ...
Страница 236: ...Ordering Information Ordering Information ...
Страница 242: ...ORDERING INFORMATION Ordering 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 243: ...Safety Data Sheets Safety Data Sheets ...
Страница 245: ...SAFETY DATA SHEETS SDS 2 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 246: ...Appendix Appendix ...
Страница 255: ......
Страница 256: ......