ThinPrep
®
3000 Processor Operator’s Manual
RECOMMENDED STAINING GUIDELINES
Staining times are different for ThinPrep-prepared slides in comparison to conventional preparations
and should be adjusted accordingly.
• Use graded concentrations of alcohol (50% or 70%) to lower the potential for osmotic shock or
possible cell shedding during staining.
• The use of mild bluing solutions and dilute acid baths will optimize nuclear staining and min-
imize possible cell shedding. Hologic recommends the use of a dilute Lithium Carbonate solu-
tion, or Ammonium Hydroxide solution as the bluing solution.
• Avoid the use of strong salt solutions, like
Scotts Tap Water Substitute
.
• Bath solution heights should completely cover the slides to reduce the chance of cell shedding
during staining.
• For optimal results, slides should be agitated for at least 10 dips in each bath.
Below are the maximum concentrations to be used for the following solutions during the staining
process:
Hydrochloric acid (HCl)
0.025%
Lithium Carbonate (bluing) baths
10 mg per 1 liter
1
Acetic acid
0.1%
Ammonium Hydroxide
0.1%
1. Refer to Bales, CE. and Durfee, GR. Cytologic Techniques in Koss, L, ed. Diagnostic Cytology
and its Histopathologic Basis. 3rd Edition. Philadelphia: JB Lippincott. Vol. II: pp 1187-1260 for
details.
SECTION
B
Содержание ThinPrep 3000
Страница 1: ......
Страница 4: ...The ThinPrep Processor The ThinPrep Processor ...
Страница 5: ...MAN 03939 001 Rev 004 page 1 of 13 Instructions for Use ...
Страница 18: ...Table of Contents Table of Contents ...
Страница 23: ...1 Introduction 1 Introduction ...
Страница 41: ...1 INTRODUCTION 1 18 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 42: ...2 ThinPrep 3000 Installation 2 ThinPrep 3000 Installation ...
Страница 54: ...2 THINPREP 3000 INSTALLATION 2 12 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 55: ...3 PreservCyt and CellFyx Solutions 3 PreservCyt and CellFyx Solutions ...
Страница 71: ...4 Sample Collection and Preparation 4 Sample Collection and Preparation ...
Страница 80: ...5 Instrument Operation 5 Instrument Operation ...
Страница 105: ...6 Maintenance 6 Maintenance ...
Страница 147: ...6 INSTRUMENT MAINTENANCE 6 42 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 148: ...7 Troubleshooting 7 Troubleshooting ...
Страница 204: ...7 TROUBLESHOOTING 7 56 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 205: ...8 Staining and Coverslipping 8 Staining and Coverslipping ...
Страница 211: ...8 STAINING AND COVERSLIPPING 8 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 212: ...9 The ThinPrep Pap Test Training Program 9 The ThinPrep Pap Test Training Program ...
Страница 215: ...10 User Interface Screens 10 User Interface Screens ...
Страница 226: ...Index Index ...
Страница 232: ...INDEX Index 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 233: ...Service Information Service Information ...
Страница 236: ...Ordering Information Ordering Information ...
Страница 242: ...ORDERING INFORMATION Ordering 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 243: ...Safety Data Sheets Safety Data Sheets ...
Страница 245: ...SAFETY DATA SHEETS SDS 2 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 246: ...Appendix Appendix ...
Страница 255: ......
Страница 256: ......