7
T
ROUBLESHOOTING
7.48
ThinPrep
®
3000 Processor Operator’s Manual
Refer to “ThinPrep Microscope Slides and Slide Cartridges” on page 5.7 for a description of the
correct slide/cartridge loading procedure.
Keep the slide cartridges free of glass chips or slide dust for optimal operation.
e. See if the fixative spray nozzle is working freely. Ordinarily a fixative clog or low CellFyx
Solution level will prompt a message specifically for that. Sometimes a clog will require fur-
ther attention to fully dissolve it. Refer to “FIXATIVE SYSTEM PREVENTATIVE MAINTE-
NANCE” on page 6.5 for detailed steps on reversing clogs. If the nozzle will not spray freely
in a few attempts, contact Hologic Technical Support.
4. Close the door(s) and press
Close
to shut the slide clamps. Press
Continue
. The instrument will
attempt recovery.
If your actions cleared the problem, the following screen will be displayed:
5. Press
Continue
. The instrument will return to the System Ready screen. Remove slides and sam-
ple trays. Place unprocessed samples into a new batch.
If the instrument diagnoses that there is still an error state, the System Fault screen will display.
Repeat steps 1 through 5, above, to try to clear any obvious mechanical jams or interference. If the
system does not recover, the following screen is displayed:
Batch Canceled Due To
Slide Handling Error
Continue
Содержание ThinPrep 3000
Страница 1: ......
Страница 4: ...The ThinPrep Processor The ThinPrep Processor ...
Страница 5: ...MAN 03939 001 Rev 004 page 1 of 13 Instructions for Use ...
Страница 18: ...Table of Contents Table of Contents ...
Страница 23: ...1 Introduction 1 Introduction ...
Страница 41: ...1 INTRODUCTION 1 18 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 42: ...2 ThinPrep 3000 Installation 2 ThinPrep 3000 Installation ...
Страница 54: ...2 THINPREP 3000 INSTALLATION 2 12 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 55: ...3 PreservCyt and CellFyx Solutions 3 PreservCyt and CellFyx Solutions ...
Страница 71: ...4 Sample Collection and Preparation 4 Sample Collection and Preparation ...
Страница 80: ...5 Instrument Operation 5 Instrument Operation ...
Страница 105: ...6 Maintenance 6 Maintenance ...
Страница 147: ...6 INSTRUMENT MAINTENANCE 6 42 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 148: ...7 Troubleshooting 7 Troubleshooting ...
Страница 204: ...7 TROUBLESHOOTING 7 56 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 205: ...8 Staining and Coverslipping 8 Staining and Coverslipping ...
Страница 211: ...8 STAINING AND COVERSLIPPING 8 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 212: ...9 The ThinPrep Pap Test Training Program 9 The ThinPrep Pap Test Training Program ...
Страница 215: ...10 User Interface Screens 10 User Interface Screens ...
Страница 226: ...Index Index ...
Страница 232: ...INDEX Index 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 233: ...Service Information Service Information ...
Страница 236: ...Ordering Information Ordering Information ...
Страница 242: ...ORDERING INFORMATION Ordering 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 243: ...Safety Data Sheets Safety Data Sheets ...
Страница 245: ...SAFETY DATA SHEETS SDS 2 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 246: ...Appendix Appendix ...
Страница 255: ......
Страница 256: ......